Omega Speaker Systems VR200 Computer Monitor User Manual


 
3 - 3
3 TRANSMITTING FUNCTIONS
3.2.2 Output Format of Measured Values in Binary Mode
When the TS0, ESC T and FM1 commands are received, the measured values and
computed results are output in binary.
Output format:
(1) Sequence of transfer
123456
Byte
Number of bytes
sent (2 bytes)
Date and time (6 bytes)
Measured data 1 (5 bytes)
Measured data n (last) (5 bytes)
(2) Number of bytes sent
Number of bytes sent = 5 × n + 6
This parameter can be sent either from the most significant byte or from the least
significant byte, as specified by the BO command.
Note that the 2 bytes of the “number of bytes sent” itself are not included.
(3) Date and time
123456
Year Month Day Hour Minute Second
Byte
Year :
Month :
Day :
Hour :
Minute :
Second :
0 to 99 (00H to 63H)
1 to 12 (01H to 0CH)
1 to 31 (01H to 1FH)
0 to 23 (00H to 17H)
0 to 59 (00H to 3BH)
0 to 59 (00H to 3BH)
3.2 Output Data Formats