4.3 Self-check
(1) Self-check Procedure
To run the self-check of the emulator, do so as explained here below. While the self-check is in
progress, the LEDs will change as shown in Figure 4.4.
(1) Set the switches in the emulation pod to the factory settings as shown in Figure 4.3. If
specified, set the switches according to your pod probe.
(2) If the target system is connected, disconnect it.
(3) Within 2 seconds of activating power to the emulator, press the system reset switch on the
emulator front panel to switch the emulator to maintenance mode.
(4) Check the SAFE LED starts blinking and then press the system reset switch again.
(5) The self-check will start. If the normal result is displayed in about 40 seconds, the self-check
has terminated normally.
IMPORTANT
Note on Self-check:
• If the self-check does not result normally (excluding target system error), the
emulation pod may be damaged. Then contact your local distributor.
Figure 4.3 Switch settings (factory-settings) for the self-check
(2) If an Error is Detected in the Self-check
If the self-check does not result normally (system status error or target status error in Figure 4.4),
check the following.
•Whether the emulation pod and PC4701 are connected properly
• Whether the proper firmware has been downloaded
• Whether the switch settings of this product are the factory settings (Figure 4.3). If specified,
check whether the switches are set according to your pod probe.