SSI America 8040 Computer Accessories User Manual


 
Super Systems Inc. Page 18 SDS Data Logger Manual #4565 Rev D
To stop a survey, highlight the Action field, enter a 0 and then press Enter. Once the survey has
finished, “Finished” will appear in the highlighted area and the stop Date/Time will be filled in the
“Stopped at” field.
To change the Setpoint, highlight the Setpoint field, enter in the new Setpoint and press Enter. Pressing
the circular arrow key while the setpoint field is highlighted will change the sign of the value entered.
The bottom half of the screen contains the full template data displayed five lines at a time: Template
Name, Company Name, Furnace ID, Furnace Make/Model, Furnace Type, Furnace Use, Furnace
Dimensions, Temperature Range, Date, Setpoint, Number of TCs, TC Type, TC Spool SN, TC Cal, TC Cal
Date, TC Spool Correction, SDS Calibrated By, SDS Cal Date, Duration, Operator, Interval, Specification,
TC Units, Simulated Load, Tolerance, Next Survey, Cont MFG, Cont Model, Cycle Time, Dead Band,
Output Limit, Prop Band, Rate, Reset, TC Gauge, and Notes. Pressing the circular arrow key will scroll
down through the template data, one line at a time. When the end of the template is reached, the scroll
will start over at the beginning of the template.
3. Display Input Settings
The Display Input Settings shows the input settings for all five thermocouples on all four/eight boards.
Pressing any of the number keys (1-9) will display the corresponding menu option. Pressing the down
arrow key will display the
Modify Input Settings
screen (menu option 8).