SSI America 8040 Computer Accessories User Manual


 
Super Systems Inc. Page 45 SDS Data Logger Manual #4565 Rev D
Figure 2.12 Define Control TC screen
Clicking on the Accept button will set the selected control TC(s). Once a control TC(s) has been selected,
the TC’s number will be in Red on the “Active Thermocouples” section of the Thermocouples tab.
The
thermocouple(s) that is identified as the control TC will not be used for the TC summary when identifying
hottest and coldest channels.
Clicking on the “View Offsets used in Survey” link will display a list of the offsets that were used on the
survey (figure 2.13).
All values that are displayed to the users are corrected values.
That means
the Offset defined for these TCs have already been incorporated into the value the operator would see.
IF there is a value for the TC Spool Correction Factor, that value will also be incorporated providing a
corrected view for the user.
Figure 2.13 TCOffsets screen