Teledyne 460H Computer Monitor User Manual


 
1.4 Specifications
Measurement Principle UV Absorption (Beer Lambert Law)
Ranges 0-5%, 0-10%, 0-15% w/w, 0-20% w/w
0-100 g/Nm
3
, 0-200 g/Nm
3
, 0-300 g/Nm
3
, 0-
400g/ Nm
3
Measurement Units wt%, g/Nm
3
Accuracy
± 1% of Full Scale
Precision/Repeatability
±0.5% of Full Scale Range
Resolution 0.1 % w/w or .01 g/Nm
3
Response Time (95%) <5 sec to 95%
Compensation Pressure, Temperature (NTP = 273.15K,
760mmHg)
Gas Inlet Pressure Range 3.0 – 30.0 psig
Gas Flow Rate 0.2 – 2.0 LPM
Temperature Range 5-45
o
C
Dimensions (H x W x D) 12.64" x 11.19" x 6.08"
(321mm x 284mm x 154mm)
Weight 9.40lb (4.27kg)
Power 110-240V~, 50/60Hz, 2.5A
Environmental Conditions Installation Category (Overvoltage Category) II
Pollution Degree 2
Maximum Operating Altitude 2000 meters
Analog Output Voltage 0-5V, 4-20mA (Optional)
Isolated Analog Output 4-20mA
Mode*
Maximum voltage between outputs and ground
60V peak
Degree of Protection (IP Code) IPX65
*Optional
P/N 03662D Teledyne API Model 460H O
3
Monitor Operator Manual - Page 10