5.4.7 TLIST
Syntax
<address>TLIST<CR>
Description
Returns a verbose list of the current measured parameters with labels for each parameter:
O3 Concentration, Cell Pressure(psia,) Cell Temperature (K,) Lamp Temperature (K,)
Measure Detector (mV,) Calibrated Reference Detector (mV,) Reference Detector (mV)
Data Parameters Sent
None
Response
Test Parameter List (See Below)
Example
Command:
1TLIST<CR>
Response:
O3 = 0.0226168
Press = 14.7753
Cell Temp = 300.7116
Lamp Temp = 324.7965
Ref = 2881.52
Meas = 2941.092
Raw Ref = 4412.646
Notes
Note that the concentration value returned shows more digits after the decimal than the
front panel display. Concentration values should only be considered valid to 4 significant
digits.
Also, note that no checksum is sent in the response to the TLIST command.