5.4.3 DACSTEP
Syntax
<address>DACSTEP<CR>
Description
Puts instrument in the Analog Output setup mode. The Analog Output steps from zero to
full-scale in 20% increments, pausing for 4 seconds at each level. This repeats 5 times,
after which the instrument returns to normal operation. This mode is useful for testing
the Analog Output and the operation of any equipment measuring the Analog Output.
Data Parameters Sent
None
Response
<address>:<success_flag>#<checksum><CR>
Example
Command:
1DACSTEP<CR>
Response:
1:OK#261<CR> - Calibration Successful
1:FAIL#391<CR> - Calibration Failed
Notes
Note that the DACSTEP function takes some time to complete. When the command is
sent to the monitor, it will immediately respond with the <address> and colon ‘:’ as an
acknowledgement that the message was received. After the function is complete the rest
of the response will be sent. No additional commands should be issued to the monitor
until the function completes.