VXI CT-100C Series Six-Slot VXIBus Chassis Network Hardware User Manual


 
VXI Technology, Inc.
24 CT-100C Installation
FAN SPEED SWITCH
A fan speed selector switch is located at the rear of the CT-100C chassis. To set the fans to
operate at HIGH speed, move the switch at the rear of the chassis to the left. At HIGH speed, the
chassis is provided maximum cooling to the instruments.
The fans will operate in the LOW speed mode when the fan speed switch is toggled to the right.
The low speed mode provides for quieter operation, but decreases the cooling capability of the
fans. To best determine which mode of operation is best in a given application, refer to
Figure 1-6
on page
19 that delineates the cooling requirements of populated mainframes. If all instruments
installed in the chassis receive adequate cooling in the low speed mode, then this mode may be
used.
VXI Technology
HIGH / LOW
WARNING - TO PREVENT POSSIBLE ELECTRIC SHOCK HAZARD, DISCONNECT POWER
CORD BEFORE REMOVING THE POWER SU PPLY MODULE FROM THE MAINFRAME.
ATTENTION - POUR EMP Ê CHER LE RISQUE POSSIBLE DE D É CHARGE É LECTRIQUE,
DEBRANCHER LE CORDON DE SECTEUR AVANT D'ENLEVER LE MODULE
D'AL IMENTATION D' É NERGIE DE L'UNIT É CENTRALE.
WARNING - FOR PROTECTION FROM ELECTRIC SHOCK, THE GROUNDING CONNECTION
IN TH E POWER CORD MUST BE PROPERLY CONNECTED.
ATTEN TION - POUR LA PROTE CTION CONTRE LE CHOC É LECTRIQUE , LA PRISE DE
TER RE AU SOL DANS LE C ORDON DE SECTEUR DOIT Ê TRE CORRECTMENT RELI É E.
WARNING - NO OPERATOR SERVICEABLE PARTS INSIDE. REFER SERVICING TO
TRAINED SERVICE PERSONNEL.
ATTENTION - AUCUNES PI È CES REPARABL ES PAR UN OP É RATEUR À L'INT É RI EUR.
R É F É REZ L'E NTRE TIEN AU PE RSONNE L QUALI FI É.
!
!
!
Fan Speed Switch Location
(Rear Panel of CT-100C Mainframe)
100-240 VAC
10A MAX.
47 - 63 Hz
100-240 VAC
10A MAX.
47 - 63 Hz
J200
FAN SPEED
HIGH / LOW
J201
MONITOR / CONTROL
FIGURE 1-3: FAN SPEED SWITCH LOCATION