Weight class and recommended measuring range
6-2-3. When Building into a System
z When a special weighing pan is to be designed, the weight of the pan and the
material to be weighed should not exceed the weighing capacity of the balance.
To reduce influences of static electricity and magnetism, use materials other than
resin and magnetic material such as iron.
z When a special weighing pan is to be designed, the weight of the pan and the
material to be weighed should not exceed the weighing capacity of the balance.
To reduce influences of static electricity and magnetism, use materials other than
resin and magnetic material such as iron.
z There is a function available to maintain the previous weight value in non-volatile
memory, even if the AC adapter is removed.
When “Zero upon power-on ( p-tr )” of “Environment, Display ( ba5fnc )” is set to
“ 1 ”, the previous weight value is displayed upon power-on.
For details, refer to “Zero upon power-on” on page 32.
z There is a function available to perform span calibration only, when performing
calibration with a tare on the weighing pan.
When “Span calibration ( 5pn )” of “Environment, Display ( ba5fnc )” is set to “ 1 ”,
span calibration using the internal mass is performed, with a tare on the weighing pan.
For details, refer to “Span calibration” on page 32.
z To set a higher response rate (weighing speed) or to batch-weigh small amounts of
material, such as a powdery material, refer to “14. Extended Function”.
Weight
(Displayed value)
MC-10K MC-30K
Model
Class
20 kg
10 kg
5 kg
2 kg
1 kg
500 g
200 g
100 g
50 g
F1 F2 M1 M2 F1 F2 M1 M2
F1 F2 M1 M2