55
Display Error code
Description
Automatic response adjustment zero error
The automatic response adjustment can not be performed
because there is something on the pan. Clear the pan. Press
the CAL key to return to the weighing mode.
(Check NG)
Automatic response adjustment unstable error
The automatic response adjustment can not be performed
because the weight value is unstable.
Check the ambient conditions such as breeze, vibration
and magnetic fields, also check the weighing pan.
Press the CAL key to return to the weighing mode.
(Check no)
Internal error
This error indicates an internal error as the result of the self-check
function. Repair is required. Contact the local A&D dealer.
Clock battery error
The clock backup battery has been depleted.
Press any key and set the time and date. The clock and
calendar function works normally as long as the AC adapter
is connected to the balance. If this error appears frequently,
contact the local A&D dealer.
(Blinking)
Memory full
The amount of weighing data in memory has reached the
maximum capacity. Delete the data in memory to store new
data. For details, refer to the GX-K manual "12. Data Memory".
(Illuminated)
Memory full
The amount of calibration or calibration test data in memory
has reached the maximum capacity (50 sets).
The data in memory will be deleted automatically to store
new data. For details, refer to the GX-K manual "12. Data
Memory".
Memory type error
The type of memory set in the function table and the type of
data stored are different. For details, refer to the GX-K manual
"12. Data Memory".
EC,E00
Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC,E01
Undefined command error
An undefined command was received.
Confirm the command.
EC,E02
Not ready
A received command can not be processed.
Example:
The balance received a "Q" command, but not in the
weighing mode.
The balance received a "Q" command while processing a
RE-ZERO command.
Adjust the delay time to transmit a command.