A&D MC-30K Postal Equipment User Manual


 
30
Class Item and Parameter Description
0
No output
d-no
Data number output
1
Output
Refer to the GX-K manual "12.
Data Memory"
0
No output
1
Time only
2
Date only
5-td
Time/Date output
3
Time and date
Selects whether or not the time
or date is added to the weighing
data. Refer to the GX-K manual
"10-7. Clock and Calendar
Function" for details.
0
No output
5-id
ID number output
1
Output
Selects whether or not the ID
number is output.
0
No pause
pU5e
Data output pause
1
Pause (1.6 seconds)
Selects the data output interval.
0
Not used
at-f
Auto feed
1
Used
Selects whether or not automatic
feed is performed.
0
No output
1
AD-8121 format
info
GLP output
2
General data format
Selects GLP output method.
For how to set time and date to
be added, refer to the GX-K
manual "10-7. Clock and
Calendar Function".
0
Not used
dout
Data output
ar-d
Zero after output
1
Used
Adjusts zero automatically after
data is output
0
600 bps
1
1200 bps
2
2400 bps
3
4800 bps
4
9600 bps
bp5
Baud rate
5
19200 bps
0
7 bits, even
1
7 bits, odd
btpr
Data bit, parity bit
2
8 bits, none
0
CR LF
Crlf
Terminator
1
CR
CR: ASCII code 0Dh
LF: ASCII code 0Ah
0
A&D standard format
1
DP format
2
KF format
3
MT format
4
NU format
type
Data format
5
CSV format
Refer to "10-6. Description of the
Item "Data Format"”.
0
No limit
t-Up
Timeout
1
1 second
Selects the wait time to receive a
command.
0
No output
erCd
AK, Error code
1
Output
AK: ASCII code 06h
0
Not used
5if
Serial
interface
Ct5
CTS, RTS control
1
Used
Controls CTS and RTS.
mlt
Programmable-unit (Multi-unit)
Available only when programmable-unit mode is selected.
Refer to the GX-K manual "14. Programmable Units" for details.
:
Factory settings. Digit is a unit of minimum weighing value.