A&D MC-30K Postal Equipment User Manual


 
32
10-4. Description of the Class "Environment, Display"
Condition ( Cond )
Cond 0 This parameter is for sensitive response to the fluctuation of a weight value. Used
for powder target weighing, weighing a very light sample or when quick response
weighing is required. After setting, the balance displays FAST.
Cond 2 This parameter is for stable weighing with slow response. Used to prevent a
weight value from drifting due to vibration or drafts. After setting, the balance
displays SLOW.
Notes In automatic response adjustment, the weighing speed is selected
automatically.
With "Hold function (Hold)" set to "ON (1)", this item is used to set the
averaging time.
Stability band width ( 5t-b )
This item controls the width to regard a weight value as a stable value. When the fluctuation
per second is less than the parameter, the balance displays the stabilization indicator and
outputs or stores the data. The parameter influences the "Auto print mode"
5t-b 0 This parameter is used for sensitive response of the stabilization indicator. Used
for exact weighing.
5t-b 2 This parameter ignores slight fluctuation of a weight value. Used to prevent a
weight value from drifting due to vibration or drafts.
Note With "Hold function (Hold)" set to "ON (1)", this item is used to set the
stabilization range.
Zero upon power-on ( p-tr )
When a hopper is attached to the weighing pan and loss-in weighing is performed, the
remaining amount of the material will become unknown if tare is performed each time a
weighing starts.
When " p-tr " is set to “ 1 ”, tare is not performed at weighing start. So, the remaining amount of
the material can be monitored, when the power is turned on again after it was turned off.
Span calibration ( 5pn )
When a hopper is attached to the weighing pan and calibration is to be performed with the
hopper attached, set " 5pn " to “ 1 ”. When the tare value (hopper and other devices attached) is
within the value in the table below, calibration using the internal mass is possible.
Model Tare value
MC-10K
9 kg or less
MC-30K
29 kg or less