11. ID Number And GLP Report
The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used.
The ID number is maintained in non-volatile memory even if the AC adapter is removed.
The GLP output format is selected at "GLP output (info)" of the function table and can be
output to a personal computer or printer using the RS-232C serial interface.
The GLP output format includes the balance manufacturer, model, serial number, ID
number, date, time and space for signature for weighing data, the weight used and results
for calibration or calibration test data.
The balance can output the following reports for GLP.
"Calibration report" of the calibration, using the internal mass (Calibration due to changes
in temperature and one-touch calibration.)
"Calibration report" of the calibration, using an external weight.
"Calibration test report" of the calibration test, using an external weight.
"Title block" and "End block" for the weighing data.
Calibration and calibration test data can be stored in memory to output several reports at the
same time. Refer to the GX-K series instruction manual "12. Data Memory" for details.
For details on confirming and setting the time and date, refer to the GX-K series instruction
manual "10-7. Clock and Calendar Function".
11-1. Setting the ID Number
Step 1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed,
then release the key.
Step 2 Press the SAMPLE key several times to display id .
Step 3 Press the PRINT key. Set the ID number using the following keys.
RE-ZERO key
.....
To set the character of the digit selected.
Refer to the display character set shown below.
SAMPLE key
......
To select the digit to change the value.
PRINT key
.........
To store the new ID number and display ba5fnc .
CAL key
............
To cancel the new ID number and display ba5fnc .
Step 4 With
ba5fnc displayed, press the CAL key to return to the weighing mode.
Display character set