Agilent Technologies N5980A Webcam User Manual


 
Programming Reference
Gating Time
:SENS1:GATE:PER
Syntax :SENS1:GATE:PER?
:SENS1:GATE:PER integer
Description Either gets or sets the duration of the gating interval. The
allowed range for the gating time is 1s to 608399s (168
hours, 59 minutes, 59 seconds). The gating time is always
in seconds. Fractional seconds are not supported, and will
be rounded to the closest time in seconds.
Gating Bit Error Rate
:SENS1:FETC:ERAT
Syntax :SENS1:FETC:ERAT?
Description Gets the bit error ration that is captured during the gating
interval. This value might/will change while the gating is in
progress, and will freeze at the end of the gating interval.
Gating Number of Errors
:SENS1:FETC:ECO
Syntax :SENS1:FETC:ECO?
Description Gets the number of errors that have been counted during
the gating interval. This value might/will change while
gating is in progress, and will freeze at the end of the
gating interval.
Start Gating
:SENS1:GATE:STAT 1
Syntax :SENS1:GATE:STAT 1|ON
Description Starts a new gating period. This command will be ignored if
a gating period is already in progress.
Abort Gating
:SENS1:GATE:STAT 0
Syntax :SENS1:GATE:STAT 0|OFF
Description This command aborts the currently active gating period
immediately. It will be ignored if the gating is not in
progress.
40 N5980A User Guide