Agilent Technologies U1251A Electronic Accessory User Manual


 
Specifications 8
Agilent U1251A/U1252A User’s and Service Guide 159
For current measurements
Peak Hold Specifications
Table 7-13 Frequency sensitivity specifications for current measurements
Input Range
Minimum Sensitivity (R.M.S. sine wave)
20 Hz – 20 kHz
500.00 µA 100 µA
5000.0 µA 250 µA
50.000 mA 10 mA
440.00 mA 25 mA
5.0000 A 1 A
Table 7-14 Peak hold specifications for dc voltage and current measurements
Signal Width Accuracy For DC mV/voltage/current
Single event > 1 ms 2% + 400 for all ranges
Repetitive > 250 µs 2% + 1000 for all ranges