Agilent Technologies U1251A Electronic Accessory User Manual


 
Specifications 8
Agilent U1251A/U1252A User’s and Service Guide 161
Square Wave Output for U1252A
Table 7-17 Square wave output specifications
Output
[1]
Range Accuracy
Frequency
0.5, 1, 2, 5, 10, 15, 20, 25, 30, 40, 50, 60,
75, 80, 100, 120, 150, 200, 240, 300,
400, 480, 600, 800, 1200, 1600, 2400,
4800 Hz
0.005% x output frequency + 2 counts
Duty Cycle
[2][4][5]
0.39% 99.60% ± 0.398% of full scale
[3]
Pulse Width
[2][4][6]
1/Frequency 0.2 ms + (range/256)
Amplitude Fixed 0 to + 2.8 V ± 0.2 V
Notes for square wave output specifications:
1 Output impedance: 3.5 kW maximum.
2 The positive or negative pulse width must be greater than 50 µs for adjusting the duty cycle or the pulse width under
different frequencies. Otherwise the accuracy and the range will be different from the definition.
3 For signal frequencies greater than 1 kHz, add 0.1% per kHz to the accuracy.
4 The accuracy for duty cycle and pulse width is based on a 5 V square wave input without dividing signal.
5 Duty cycle can be set for 256 steps and each step is 0.390625% per kHz.
6 The pulse width can be set for 256 steps and each step is 1/(256 x Frequency).