1) Fail Counter (higher byte)
2) Fail Counter (lower byte)
3) 255 (FFh) Operation Complete Byte
or
1) 255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Module not attached
238 (EEh) Time Out Error
Clear Module Fail Counter – Control Word (A205h) (Option 6 only)
Description: Sets the module lock fail counter to 0.
Bytes to Follow: 0 bytes
Site Master Returns: 1 byte
1) 255 (FFh) Operation Complete Byte
224 (E0h) Parameter Error: Module not attached
238 (EEh) Time Out Error
Remote Self Test - Control Word (AA15h)
Description: Trigger the equivalent of a "key press" selftest.
NOTE: The response bytes will not all be returned immediately. The first 12 will be returned, then there
will be a slight delay before the next 14 are returned, then a final delay while the T1/E1 selftest is per-
formed and the final 12 bytes are returned.
Bytes to Follow: 0 bytes
Site Master Returns:
S331D (No Options/Option 3): 25 bytes
1) Temperature in 1/10th of degree Celsius (e.g., 362 = 36.2° C) (higher byte)
2) Temperature in 1/10th of degree Celsius (e.g., 362 = 36.2° C) (lower byte)
3) Memory Check (01h: Pass, 00h: Fail)
4) RTC Voltage Check (01h: Pass, 00h: Fail)
5) Power Voltage in 1/10ths of a Volt (e.g., 124 = 12.4 Volts) (higher byte)
6) Power Voltage in 1/10ths of a Volt (e.g., 124 = 12.4 Volts) (lower byte)
7) VNA Lock Failure Counter (higher byte)
8) VNA Lock Failure Counter (lower byte)
9) VNA Integrator Failure Counter (higher byte)
10) VNA Integrator Failure Counter (lower byte)
11) SPA LO Failure Counter (higher byte)
12) SPA LO Failure Counter (lower byte)
13) H/W Config - Mother Board ID
14) H/W Config - SPA Board ID
15) H/W Config - T1E1 Board ID
16) H/W Config - PLD1 ID
17) H/W Config - PLD2 ID
18) H/W Config - T1E1 COLD ID
19) VNA PLL Lock Failure Test - Status (01h: Pass, 00h: Fail)
20) VNA PLL Lock Failure Test - Failed data point # (Ignore this byte if the Lock Fail Test Status was Pass)
21) VNA PLL Lock Failure Test - Failed PLL # (Ignore this byte if the Lock Fail Test Status was Pass)
22) VNA Integration Test - Status (01h: Pass, 00h: Fail)
23) VNA Integration Test - Failed data point # (Ignore this byte if the Integration Test Status was Pass)
Site Master PM 105