Barco E1 Computer Monitor User Manual


 
NioWatch operation
System manual E1 79
Figure 40: Match with reference
To equalize the display, check the option “Match this display
with display....” and click Next. The display number that
appears in this option, refers to the reference display.
Manual calibration
1. Follow step 2 and 3 of the calibration with Barco LCD sensor
(see page 77).
2. In the Calibration dialog (see step no. 4. in the description of
the calibration with sensor), check the option Visually and
click the OK button.
The DICOM Optimizer dialog appears.