Barco E1 Computer Monitor User Manual


 
NioWatch operation
80 System manual E1
Figure 41: DICOM Optimizer dialog
3. Adjust the slider until there is almost no visible difference
between the background of the image above the slider and
the bitmap inside this image.
4. When done, click button Set point 1 of 20.
5. Repeat this procedure until the button is grayed. This
indicates the last point is set.
6. Click OK to finish.
To reset to the latest calibration:
1. Follow step 2 and 3 of the calibration with Barco LCD sensor
(see page 77).
2. In the Calibration dialog (see step no. 4. in the description of
the calibration with sensor), check the option Use the latest
calibration.