Brady 2461 Printer User Manual


 
65
5.0.2 Advanced Entry Calibration
Advanced Entry Calibration is the alternate method for special-case media
types. The procedure has two parts:
1) Using different sampling algorithms, displayed sensor readings have to be
listed for each label and TOF value.
2) From this data, the best algorithm is chosen; then those values are manually
entered for the ‘Paper’, ‘Gap’ (or ‘Mark’), and ‘Empty’ variables.
+
Advanced Entry Calibration will override all previous calibration
settings; this method should be used only when Standard Calibration
has failed.
To perform an Advanced Entry Calibration:
Step Operator Action Displayed Message Comment
1
(Media should be loaded
and the appropriate
Sensor Type selected.)
Turn ‘On’ the printer.
Wait until the
printer initializes
(about six
seconds).
2
Press the key.
Raise the printhead
assembly.
You are entering
Menu Mode.
3
Press the key to
enter the Media Settings
menu.
See Section 4.1.2
for menu layout.
4
Press the key
and scroll to ‘Sensor
Calibration’.
Press the
key to abort this
procedure.
5
Press the
key.
You are entering
the calibration
submenu.