Braun MCA166-USB Network Card User Manual


 
MCA166-USB
3.13 Miscellaneous
For communication and evaluation some libraries for MS Windows and Linux are
available. This may help the make own application programs or to integrate the
MCA166-USB in an existing system.
4 MCA166-USB parameters for use with different detectors
This table shall just give an overview of possible settings useful with the MCA166-USB. It
is far from being complete. For actual settings with your detector, see the detectors
manual.
Table 2: Parameters for different detectors
Type
Model
Manufact.
HV (V)
+ 12 V
-12V
+ 24V
- 24 V
Input signal
polarity
Channels
Gain for 2
MeV
full range
Shaping
time (µs)
CdZnTe SDP/Z/60 Ritec +300 x x - - neg 1024 90 1
CdZnTe SDP/Z/20 Ritec +500 x x - - neg 1024 80 1
CdZnTe CZT500 Ritec +600 x x - - pos 1024 128 1
NaI Scintiblock Crismatec +500 pos 512 7.2 1
NaI 25825 Scionix +550 x x - - pos 512 90 1
NaI 40*40 Amcrys-h -800 x x pos 512 14 1
CsI Sc1010 Eurorad - x x - - neg 256 160 1
HPGe GMX-20190-S Ortec -3000 x x x x neg 4096 18 2
HPGe GEM 15-190 Ortec +3000 x x x x pos 4096 24 2
HPGe GR 2018 Canberra -3000 neg 4096 11 2
HPGe GL 0310 Canberra -2000 x x x - neg 4096 2.3 2
Si
planar
7865S Ortec -1000 neg 4096 1.9 2
HPGe EGC 30-190R Eurisys -3000 pos 4096 52 2
HPGe EGPC 25-185 Eurisys +2000 neg 4096 35 2
14