Braun MCA166-USB Network Card User Manual


 
MCA166-USB
MCA Stabilisation Commands:
Command name
CMD_SET_STABILISATION
Description
The command sets the peak stabilisation.
fl: 0 turns stabilisation off
1 turns stabilisation on
rb+3<fl<re-3 stabilisation to channel fl
rb: Peak ROI begin (LLD<= begin < end)
re: Peak ROI end (begin < end <= ULD, (end-begin) < 250)
Command syntax
Format integer integer integer long integer
Parameter preamble command fl rb re end flag
Byte String (HEX) A5 5A 4D 00 fl
l
fl
h
rb
l
rb
h
re
l
re
h
B9 9B
Used setup value channels, lld, uld
Affected setup value stab_state, stab_roi_begin, stab_roi_end
MCA state STATE_READY
Command name
CMD_SET_STAB_PARAM
Description The command sets the stabilisation parameters.
st: time interval (default 10)
sa: area (default 25000)
Command syntax
Format integer integer integer long integer
Parameter Preamble command st sa end flag
Byte String (HEX) A5 5A 67 00 st
l
st
h
sa
l
sa
l
sa
l
sa
h
B9 9B
Used setup value -
Affected setup value stab_time, stab_area
MCA state STATE_READY
66