Braun MCA166-USB Network Card User Manual


 
MCA166-USB
Command name
CMD_QUERY_USER_DATA
Description
The command reads a 32 bit value from the MCA parameter memory.
e: 0 ... 63parameter number
Command syntax
Format integer integer integer long integer
Parameter preamble command e 0 end flag
Byte String (HEX) A5 5A 5E 00 e 00 00 00 00 00 B9 9B
Used setup value None
Result see MMCA_QUERY_USER_DATA function in DLL description.
Command name CMD_SYSTEM_DATA
Description The command reads the special MCA data.
Command syntax
Format integer integer integer long integer
Parameter preamble command 00 End end flag
Byte String (HEX) A5 5A 62 00 00 00 00 00 00 00 B9 9B
Used setup value
Result see MMCA_QUERY_SYSTEM_DATA function in DLL description.
Command name
CMD_QUERY_CENTROID
Description The command reads the centroid of a ROI (beg<end, LLD<=beg, end<=ULD, end-beg<250).
Command syntax
Format integer integer integer long integer
Parameter preamble command beg End end flag
Byte String (HEX) A5 5A 5F 00 beg
l
beg
h
end
l
end
h
00 00 B9 9B
Used setup value channels, lld, uld
Result see MMCA_QUERY_CENTROID function in DLL description.
70