Braun MCA166-USB Network Card User Manual


 
Command name
CMD_SET_PZC
Description
The command performs the Pole Zero Cancellation.
pv: 0 ... 2499 PZC control voltage
pm = 0 set PCZ only
pm > 0 set PZC and measure PZC offset
Command syntax
Format integer integer integer long integer
Parameter preamble command pm Pv end flag
Byte String (HEX) A5 5A 58 00 pm
l
pm
h
pv
l
pv
h
00 00 B9 9B
Used setup value None
Affected setup value pzc_value
Results If pm > 0 the function returns with the number of counts and the - PZC offset.
MCA state STATE_READY
Command name CMD_SET_PZC_TIME_OFFSET
Description The command sets the time offset for Pole Zero Cancellation.
t1: 0 ... 31 Time offset for low shaping time
t2: 0 ... 31 Time offset for high shaping time
Command syntax
Format integer integer integer long integer
Parameter preamble command t1 t2 end flag
Byte String (HEX) A5 5A 60 00 t1 00 t2 00 00 00 B9 9B
Used setup value None
Affected setup value pzc_dtc_1_offset, pzc_dtc_3_offset
MCA state STATE_READY
65