Fluke 80 Series III Power Supply User Manual


 
Specifications
47
Table 15. Capacitance and Diode Function Specifications
Function Range Resolution Accuracy
1
E
5.00 nF
0.0500 µF
0.500 µF
5.00 µF
0.01 nF
0.0001 µF
0.001 µF
0.01 µF
±(1% + 3)
±(1% + 3)
±(1% + 3)
±(1.9% + 3)
G
3.000 V 0.001 V ±(2% + 1)
1. With a film capacitor or better, using Relative mode to zero residual. See the first sentence in Table 11 for a complete explanation of
accuracy.
Table 16. Frequency Counter Specifications
Function Range Resolution Accuracy
1
Frequency
(0.5 Hz to 200 kHz,
pulse width >2 µs)
199.99
1999.9
19.999 kHz
199.99 kHz
>200 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
±(0.005% + 1)
±(0.005% + 1)
±(0.005% + 1)
±(0.005% + 1)
unspecified
1. See the first sentence in Table 11 for a complete explanation of accuracy.