Kodak i260 Scanner User Manual


 
A-61167 September 2002 D-5
Calibration begins. A message appears when the ADF calibration
has finished.
6. Raise the scanner’s input tray and rest it against the scanner.
7. Lift the flatbed cover and hold it up.
8. Place the calibration target face down on the glass platen.
9. Position the calibration target with the corner aligned with the arrow.
10. Close the flatbed cover.
11. Click OK.
Calibration begins. A message appears when the flatbed calibration
has finished.
12. Click OK.
Cleaning the Flatbed The flatbed’s glass platen can collect fingerprints and dust which may lessen
the quality of your scans. Use Staticide Wipes for KODAK Scanners (Catalog
No. 896-5519) to clean the platen glass.
1. Open the flatbed cover.
2. Wipe the platen glass with a Staticide Wipe.
3. Dry the platen glass with a lint-free cloth.
4. Close the platen cover.
After a Power Failure If power is interrupted while you are using the flatbed, the scanner camera
may not be in its home position. Do not remove the flatbed if this has occurred.
Once power has been restored, power up the scanner so that the camera can
return to its home position.