35
Field Calibration
2Turn the lock nut counterclockwise to unlock the
rubber foot block.
M
emo
Turn clockwise for lower heights and counterclockwise for
higher heights.
Note
Because there is no preventive structure, moving the rub-
ber foot block more than adjustable range may cause the
rubber foot block to fall off. In this case, screw the rubber
foot block onto the stopper unit clockwise.
3After adjusting the height correctly, move the lock
nut to the stopper unit by turning it clockwise and
lock the rubber foot block.
3
Attach the Unit Attachment Section to the
Field Calibration System Unit Section.
Use the 4 screws and tighten them firmly with the in-
cluded hex wrench.
4
Attach the VIVID 9i Attachment Section to the
main unit.
Make sure the footings at the base of the main unit are
inserted into the 4 footing seats.
Adjust the height
by turning the Rub-
ber foot block.
• The lock nut
moves in the
same height di-
rection as the
rubber foot block.
After adjustment of
the height, move
the lock nut to the
stopper unit by
turning it clockwise
and lock the rub-
ber foot block in
place.
Vivid910-hard-E.book Page 35 Monday, October 18, 2004 3:11 PM