40
Field Calibration
Ending Field Calibration
After the field calibration is finished, remove the unit from the Field Calibration System. Also close the Field
Calibration System that was used and return it to its original condition.
[Procedure]
1
Remove the unit from the Attachment Section
of the main unit.
2
Close the Field Calibration System.
Note
• Be careful not to pinch your fingers in the opening and
closing section when you shut the Field Calibration Sys-
tem.
• The Field Calibration System may be degraded if left
open. Always close and store the Field Calibration Sys-
tem after use.
• In the case where the MIDDLE/TELE lens type Field
Calibration System has been opened after opening the
WIDE lens type Field Calibration System, the Field Cal-
ibration System for the WIDE lens type and the Unit Sec-
tion Cover will be separated so be sure to close the Unit
Section Cover only after closing the WIDE lens type
Field Calibration System.
The Field Calibration System can be stored as is, without disassembling it into the Field Calibration System
Unit Section and the Frame Section.
In the case where the system has to be disassembled and stored for reasons related to installation space require-
ments or to move or transport it, follow the procedure outlined below to disassemble it.
1Close the Field Calibration System
for the MIDDLE/TELE lens type.
2Close the Field Calibration System
for the WIDE lens type.
* In the case where the MIDDLE/TELE lens type
Field Calibration System has been opened after
opening the WIDE lens type Field Calibration Sys-
tem, first close the WIDE lens type Field Calibra-
tion System and then close the unit section cover.
Vivid910-hard-E.book Page 40 Monday, October 18, 2004 3:11 PM