Konica Minolta VI-9i Graphics Tablet User Manual


 
36
Field Calibration
Field Calibration System Preparation
There are two types of field calibration systems: one for WIDE lens type use and the other for MIDDLE/TELE
lens type use. Use the Field Calibration System that corresponds to the type of lens the main unit is equipped
with.
[Procedure]
Preparing the Wide Lens Field Calibration System
1
Open the cover of the Field Calibration Sys-
tem Unit Section.
1Pull Lock Lever A down to the UNLOCK posi-
tion.
2Grasp Handle A and Handle B with both hands
and pull to spread open.
Note
There may be instances where the Field Calibration Sys-
tem for the WIDE lens will slightly jut out when the Unit
Section Cover is opened.
When using the Field Calibration System for the WIDE
lens, gently push on it to return the Field Calibration Sys-
tem to the unit side.
Preparing the MIDDLE/TELE Lens Field Calibration Sys-
tem
1
Open the Field Calibration System Unit for the
WIDE lens.
1Pull Lock Lever B down to the UNLOCK posi-
tion.
2Grasp Handle B and the Support Section of the
Field Calibration System Section using both
hands and pull to spread open.
M
emo
The Field Calibration System for the MIDDLE/TELE lens
may be opened using the procedure for opening the Field
Calibration System unit for the WIDE lens or it may be
opened by opening the Unit Section Cover and then opening
the Field Calibration System for the WIDE lens.
1
2
Pull Lock Lever A down to
the UNLOCK position.
Grasp firmly with
both hands and pull
to open.
* There may be cases in which
the Field Calibration System
for the WIDE lens type will jut
out slightly when the Unit
Section Cover is opened, so
gently push on it to return it
to its position.
1
2
Pull Lock Lever B down
to the UNLOCK posi-
tion.
Grasp firmly with
both hands and pull
to open.