Appendix C Error Codes and Solutions
© National Instruments Corporation C-3 NI-488.2 User Manual for Windows
•
ibcac
and
ibgts
.
• Any NI-488.2 multi-device calls that issue GPIB command bytes:
SendCmds
,
PPoll
,
Send
, and
Receive
.
Solutions
Possible solutions for this error are as follows:
• Use
ibsic
or
SendIFC
to make the GPIB interface become CIC on
the GPIB.
• Use
ibrsc
1
to make sure your GPIB interface is configured as
System Controller.
• In multiple CIC situations, always be certain that the CIC bit appears
in the status word
ibsta
before attempting these calls. If it does not
appear, you can perform an
ibwait
(for CIC) call to delay further
processing until control is passed to the interface.
ENOL (2)
ENOL usually occurs when a write operation is attempted with no
Listeners addressed. For a device write, ENOL indicates that the GPIB
address configured for that device in the software does not match the GPIB
address of any device connected to the bus, that the GPIB cable is not
connected to the device, or that the device is not powered on.
ENOL can occur in situations where the GPIB interface is not the CIC and
the Controller asserts ATN before the write call in progress has ended.
Solutions
Possible solutions for this error are as follows:
• Make sure that the GPIB address of your device matches the GPIB
address of the device to which you want to write data.
• Use the appropriate hex code in
ibcmd
to address your device.
• Check your cable connections and make sure at least two-thirds of
your devices are powered on.
• Call
ibpad
(or
ibsad
, if necessary) to match the configured address
to the device switch settings.