NEC MD213MC Computer Monitor User Manual


 
English-1
English
WARNING
CAUTION
TO REDUCE THE RISK OF ELECTRIC SHOCK, MAKE SURE POWER CORD IS UNPLUGGED FROM WALL SOCKET.
TO FULLY DISENGAGE THE POWER TO THE UNIT, PLEASE DISCONNECT THE POWER CORD FROM THE AC
OUTLET. DO NOT REMOVE COVER (OR BACK). NO USER SERVICEABLE PARTS INSIDE. REFER SERVICING TO
QUALIFIED SERVICE PERSONNEL.
TO PREVENT FIRE OR SHOCK HAZARDS, DO NOT EXPOSE THIS UNIT TO RAIN OR MOISTURE. ALSO, DO NOT
USE THIS UNIT'S POLARIZED PLUG WITH AN EXTENSION CORD RECEPTACLE OR OTHER OUTLETS UNLESS
THE PRONGS CAN BE FULLY INSERTED.
REFRAIN FROM OPENING THE CABINET AS THERE ARE HIGH VOLTAGE COMPONENTS INSIDE. REFER
SERVICING TO QUALIFIED SERVICE PERSONNEL.
Power Cord Important Information
CAUTION: Please use the power cord provided with this display in accordance with the table below. If a power cord is not
supplied with this equipment, please contact your supplier. For all other cases, please use a power cord that matches the
AC voltage of the power outlet and has been approved by and complies with the safety standard of your particular country.
When you use this Display in North America, you should use a North America Hospital Grade power cord.
Plug Type North America
European
Continental
U.K. Chinese Japanese
Plug Shape
Country
Voltage
U.S.A./Canada U.K. China JapanEU (except U.K.)
120 230 220 100230
NOTE: This product can only be serviced in the country where it was purchased.
(green dot
and tag label)
Intended Use
MD212MC/MD213MC/MD213MG are intended to be used for displaying and viewing of digital images for diagnosis by trained
physicians.
Caution: MD212MC/MD213MC/MD213MG cannot be used for a life-support system.
This unit is designed as component of a final system which is in compliance to IEC60601-1-1 requirements.
Safety Precaution
The unit is for exclusive interconnection with IEC60XXX certified equipment outside of patient environment and IEC 60601-1
certified equipment inside the patient environment.
Equipment connected to the analog or digital interfaces must comply with the respective IEC standards (e.g. IEC 60950 for
data processing equipment and IEC 60601-1 for medical equipment).
This device complies with EN60601-1-2. To minimize the interference from other equipment, a minimum 0.5 m distance shall
be kept form other potential electromagnetic sources such as cell phones.
•To reduce the risk of electric shock, make sure power cord is unplugged from wall socket. To fully disengage the power to
the unit, please disconnect the power cord from the ac outlet. Do not remove any cover. No user serviceable parts inside.
Refer servicing to qualified service personnel. The AC outlet shall be readily available and accessible.
This device uses temperature-controlled fans for internal cooling. The use of this device in surgery rooms is not
recommended if cleaning of the air intakes from dust and textile lint in regular time intervals cannot be performed by the user
(interval: max. six months).
North America Customers
Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked “Hospital
Only” or “Hospital Grade”. That signal input port need to be connected properly and that any unused signal input port shall not
be accessible for the patient to contact in the patient area after the LCD is integrated into a medical system.
Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner.
European Customers
Unpacking, Installation and calibration of this display should only be done by authorized and trained personnel. Any
installation done by a non-authorized person is done under his own risk and we accept no responsibility in any device
malfunctioning.
Medical Imaging
MD212MC is designed for 1200 x 1600 (portrait mode) or 1600 x 1200 (landscape mode) medical imaging.
MD213MC and MD213MG are designed for 1536 x 2048 (portrait mode) or 2048 x 1536 (landscape mode) medical imaging.