INSTRUMENT CALIBRATION
2-3
Tab Down Tab Up Enter Back Read On/Off
Calibration Procedure
The calibration procedure consists of two white reference measurements followed by
four zero reflectance measurements.
1. Use Tab Down or Tab Up to highlight Calibrate. Press Enter to access the calibration
function.
–MAIN MENU–
Store
Normal
Reference
Calibrate
Config.
MA48
——————
XXXX
******
The instrument displays the calibration status and reference serial number. The serial
number displayed must match the number listed on the calibration reference. If the
calibration status is OK, no calibration is required at this time. If Cal Required is the
current status, you must perform the calibration procedure before continuing with any
additional measurements.
Calibration
Measure White Ref
Press Firmly
Status: Cal Required
S/N: XXXXX
2. Remove the white calibration reference from the docking station and position the
instrument on the reference as previously discussed. Hold the instrument down and
momentarily press one of the Read buttons. Hold steady until the instrument indicates
that both the white calibration readings are complete.
3. After successful white calibration readings, position the instrument in its normal position
on the docking station. The display indicates that the Black Ref measurements are now
required. Hold the instrument steady against the docking station and momentarily press
one of the Read buttons to take the measurements. You must keep the instrument steady
through all four measurements.
After the measurements are completed, the display automatically returns back to the main
menu.