4. Instrument Operation
Reference Mode
The instrument can store a maximum of 100 references with tolerances which are
accessible in the Store and Normal modes. The reference mode is used to measure
references and edit their tolerances. References can also be downloaded from an X-Rite
software application, such as X-RiteColor Master. If a reference is downloaded, it can be
viewed but not edited.
1. Use Tab Up or Tab Down to highlight References.
MAIN MENU
Store
Normal
References
Calibrate
Config.
MA48
——————
XXXX
******
2. Press Enter to access the reference mode.
Reference: 1
Tolerance: L
*
a
*
b
*
Edit Tolerances
31.43 24.57 262.34
30.50 24.45 262.14
32.41 26.03 262.19
Selecting a Reference Number and Measuring
The reference mode has a special control that is used to select a reference number. The
list can be paged through either forward or backward to reduce time.
Reference number will be replaced with the reference name when downloaded from
X-RiteColor Master software.
1. Use Tab Up or Tab Down to highlight Reference: ##.
2. Press Enter to activate the editor. Up/down arrows appear at the left.
Reference: 1
Tolerance: L
*
a
*
b
*
Edit Tolerances
31.43 24.57 262.34
30.50 24.45 262.14
32.41 26.03 262.19
3. Press Tab Down to page backward through the reference numbers, or press Tab Up to
page forward through the list. To quickly page through the references, hold down one of
the Tab keys.
4. Press Enter to select the displayed reference number.
NOTE: Use Back to revert back to the original reference.
5. Refer to section 1: Measurement Techniques for information on proper positioning.
Position the instrument over the reference and take the measurement. The selected angles
appear momentarily during the measurement.
NOTE: If the averaging option is activated in the Configuration mode, you must
make additional measurements on the reference. Refer to section 1:
Measurement Averaging for additional information.
Current reference number
Tolerance type for selected reference
A
ccesses tolerance editing menu
A
ngle data for current reference
Indicates the edito
r