6-27
Calibrating for Increased Measurement Accuracy
One-Port Reflection Error Correction
NOTE Include any adapters that you will have in the device measurement. That is,
connect the calibration standard to the particular connector where you will
connect your device under test.
Figure 6-8 Standard Connections for a One Port Reflection Error Correction
8. To measure the standard, when the displayed trace has settled, press:
NOTE If the calibration kit that you selected has a choice between male or female
calibration standards, remember to select the sex that applies to the test port
and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the softkey after it measures the
calibration standard.
9. Disconnect the open, and connect a short circuit to the test port.
10.To measure the standard when the displayed trace has settled, press:
The analyzer measures the short circuit and underlines the softkey.
11.Disconnect the short, and connect an impedance-matched load to the test port.
12.When the displayed trace has settled, press:
, select the type of load you are using, and then press when
the analyzer has finished measuring the load.
Notice that the softkey is now underlined.
13.To compute the error coefficients, press:
OPEN
OPEN
SHORT
SHORT
LOADS
DONE: LOADS
LOADS
DONE: 1-PORT CAL