Agilent Technologies 6622A Power Supply User Manual


 
Output Connections and Operating Information 59
Voltage All series specifications referring to voltage are twice the single output specification except for
programming resolution which is the same as for a single output.
Current All series specifications referring to current are the same as for a single output except for CC load
effect, CC load cross regulation, CC source effect, and CC short term drift which are twice the current
programming accuracy (including the percentage portion).
Load Transient
Recovery Time
Load transient recovery time is the same to within approximately twice the voltage setting band since
the output impedances of the series combination add together.