Agilent Technologies FS2100 Webcam User Manual


 
Minimum Clock Period
(State)
0 to 33Mhz PCI clock for State mode and 0-100Mhz for Timing
Mode.
Signal loading
The PCI Active Analysis Probe logic analyzer interface presents
only one electrical load on each PCI bus signal. However, the
extender card connector is an additional 4 inches beyond the
maximum allowed stub length.
Operations
All PCI Local Bus operations supported.
Environmental
Temperature
Operating:0 to 55 degrees C (+32 to +131 degrees F)
Non operating:-40 to +75 degrees C (-40 to +167 degrees F)
Altitude
Operating: 4,6000m (15,000 ft)
Non operating: 15,3000m (50,000 ft)
Up to 90% non condensing. Avoid sudden, extreme temperature
changes which would cause condensation on the Analysis Probe
module.
Humidity
There are no automatic performance tests or adjustments for the
PCI Active Analysis Probe module. If a failure is suspected in
the PCI Active Analysis Probe module contact the factory or your
FuturePlus Systems authorized distributor.
Testing and
Troubleshooting
Servicing
The repair strategy for the PCI Active Analysis Probe is module
replacement. However, if parts of the PCI Active Analysis Probe
module are damaged or lost contact the factory for a list of
replacement parts.