Agilent Technologies U2751A Switch User Manual


 
Getting Started 1
U2751A User’s and Service Guide 27
U2922A Terminal Block Installation
This section provides the recommended procedure for
connecting the U2922A terminal block to the U2751A.
WARNING
The maximum working voltage of the U2751A with the
terminal block for standalone is 35 Vrms and for
modular (when used with the U2781A) is 180 Vrms.
The maximum transient voltage is 300 Vrms.
Do not remove the retractable cover from the U2922A
terminal block during operation to avoid any
unexpected hazard.
NOTE
You are required to connect the cables to the U2922A terminal block
prior to attaching the U2922A to the U2751A.
Ensure that you power-off your device and unplug the U2922A from the
U2751A to change the cable connection on the U2922A.
Connect the cables to the terminal block as
desired.
cables