Agilent Technologies U2751A Switch User Manual


 
40 U2751A User’s and Service Guide
3 Characteristics and Specifications
Product Characteristics
REMOTE INTERFACE
1
Hi-Speed USB 2.0
USBTMC 488.2 Class device
POWER CONSUMPTION
+12 VDC, 2 A maximum
Installation Category III
OPERATING ENVIRONMENT
Operating temperature from 0 C to +50 C
Relative humidity at 20% to 85% RH (non-condensing)
Altitude up to 2000 meters
Pollution degree 2
For indoor use only
STORAGE COMPLIANCE
–20 C to +70 C
SAFETY COMPLIANCE
Certified with:
IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
Canada: CAN/CSA-C22.2 No. 61010-1-04
USA: ANSI/UL 61010-1:2004
EMC COMPLIANCE
IEC 61326-2002/EN61326:1997+A1:1998+A2:2001+A3:2003
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS CISPR11:2004
SHOCK AND VIBRATION
Tested to IEC/EN 60068-2
I/O CONNECTOR
DSub 25 male
DIMENSIONS (W D H)
105.00 175.00 25.00 mm (without bumpers)
117.00 180.00 41.00 mm (with bumpers)
WEIGHT
428 g (without bumpers)
480 g (with bumpers)
WARRANTY
One year
CALIBRATION
Annual calibration is not required
1 For remote connections using Agilent E5813A, refer to Chapter 1.