Anritsu MA24106A Network Router User Manual


 
Buttons Using the Power Sensor
3-2 MA24106A UG
3-3 Buttons
There are nine buttons available on the user interface to perform the most common tasks as described below:
•Zero: Performs the Zero operation. Removes system noise.
Hold/Run: Holds the last reading. Run releases the hold.
•Frequency: Applies frequency correction to the measured power.
•Power Units: Displays units of power in linear or log scale.
Normal/Relative: Displays power changes with respect to desired reference value.
Averages: Facilitates custom averaging number entry. The default number is 1.
•Fixed Offset: Facilitates an offset correction in dB. Input positive value for attenuation.
Apply: Applies the current entry in the data entry field.
•Exit: Terminates the program.
3-4 Data Entry Fields
The data entry fields become active when clicking a button to accept an appropriate entry (see the example in
Figure 3-2). Communication with the power sensor does not take place until the Apply button is clicked or the
Enter key on the keyboard is pressed. The following list summarizes the entry fields:
Frequency Button: Sets the Cal Factor Frequency in GHz
Power Units Button: Sets the units of power to dBm, W, mW, or µW
Averages Button: Sets the number of averages from 1 to 256
Fixed Offset Button: Sets the attenuation from –100 dB to +100 dB Fixed Offset
Figure 3-2. Example of an Active Entry Field