Anritsu MA24106A Network Router User Manual


 
Remote Operation Command Details Remote Operation
4-2 MA24106A UG
4-3 Remote Operation Command Details
Each command needs to be followed by a Line Feed (0x0A, ASCII 10) termination character. Response from the
sensor will have a Line Feed (0x0A, ASCII 10) termination character attached at the end.
START
Description: Puts the power sensor in measurement mode.
Syntax: START + LF
Return Value: None for the first time, OK for any subsequent command sent.
Remarks: This command does not return anything when sent the first time. For any subsequent
START commands, the sensor will return OK. This is helpful, if the user wants to know if
the sensor is in measurement mode.
STOP
Description: Stops the measurement mode and puts the sensor in idle mode.
Syntax: STOP + LF
Return Value: OK or ERR
Remarks: This command should be sent before exiting the user application.
IDN?
Description: Gets identification information from the sensor.
Syntax: IDN? + LF
Return Value: ANRITSU, Model #, Serial #, Module Serial #, firmware version
PWR?
Description: Gets the power reading from the power sensor.
Syntax: PWR? + LF
Return Value: Power value in dBm
Remarks: If an error condition exists, the returned power reading values are pre-tagged with the
letter “E” and the sensor’s LED turns yellow. The STATUS? command can then be issued
to find details about the error.
NPWR?
Description: Gets a new power reading from the power sensor.
Syntax: NPWR? + LF
Return Value: Current Power value in dBm
Remarks: After receiving this command, the power sensor discards the existing data that is stored
in the sensor’s buffer. A new measurement of the current power is initiated to get a new
power reading.
If an error condition exists, the returned power reading values are pre-tagged with the
letter “E” and the sensor’s LED turns yellow. The STATUS? command can then be issued
to find details about the error.
Note: This command is only available in sensor firmware versions 1.01 and later.