Linearity Test Sensor Operational Tests
5-10 MA24106A UG
Table 5-7. Measurement Results (6 GHz)
Row
#
Synthesizer
Power Level
Setting
(dBm)
Attenuation
in Reference
Arm
(dB)
A
Reference
Power
Measurement
(dBm)
B
= (A6–A5)
Correction
(dB)
C
= (A+B)
Corrected
Reference Power
Measurement
(dB)
Attenuation
in Test Arm
(dB)
D
MA24106A
Measurement
(dBm)
E
= (C–D)
Difference
Calculation
(dB)
1 +20 10 0
2 +15 10 0
3 +10 10 0
4+5 10 0
50 10 0
6 adjust per
step 7e
00 10
7+5 0 0 10
80 0 0 10
9–5 0 0 10
10 –10 0 0 10
11 Max:
12 Min:
13 Delta (E11 – E12):