Using the Power Sensor Making Measurements
MA24106A UG 3-13
Table 3-1, describes the number of averages needed to attain a certain noise level for a particular power level
measurement with the Low Aperture Time mode setting.
Table 3-2, describes the number of averages needed to attain a certain noise level for a particular power level
measurement with the High Aperture Time mode setting.
Table 3-1. MA24106A Averaging Table (Low Aperture Time, Default Mode)
Input Power
(dBm)
Input Power
(mW)
Number of
Averages
Needed for
<
±0.20 dB
Noise
Number of
Averages
Needed for
< ±0.15 dB
Noise
Number of
Averages
Needed for
< ±0.10 dB
Noise
Number of
Averages
Needed for
< ±0.05 dB
Noise
Number of
Averages
Needed for
< ±0.01 dB
Noise
2010011111
1531.611111
1010.011111
53.1611112
01.00111416
-5 0.316 1 1 1 20 78
-100.10011111
-150.031611111
-200.010011117
-250.00316111361
-30 0.00100 2 3 7 25
-35 0.000316 16 28 62 245
-400.000100158––––
Table 3-2. MA24106A Averaging Table (High Aperture Time)
Input Power
(dBm)
Input Power
(mW)
Number of
Averages
Needed for
<
±0.20 dB
Noise
Number of
Averages
Needed for
<
±0.15 dB
Noise
Number of
Averages
Needed for
<
±0.10 dB
Noise
Number of
Averages
Needed for
<
±0.05 dB
Noise
Number of
Averages
Needed for
<
±0.01 dB
Noise
2010011111
1531.611111
1010.011111
53.1611111
01.0011111
-50.31611125
-100.10011111
-150.031611111
-200.010011111
-250.0031611114
-300.00100111238
-35 0.000316 1 2 4 16
-40 0.000100 10 18 39 153