Epson 9840 Computer Monitor User Manual


 
INTERFACE MODEL 9840 VER 6.0.2
from the cell before entering this command since it will perform the shunt calibration
measurement that is recorded for use with the calibration check command. The format
is CTP0.
Example:
@123CTP0
Acknowledge:
@123 Calibrate Command - Reading for Shunt Check...
[10 seconds delay]
@123 Calibrate Command Completed
Ch A = S/N 123456, 100.00 LbI , 3.00150 mV/v,
3.00230 mV/v,
3.00238 mV/v,
3.00353 mV/v,
10.00 V , Cal on Oct27-99, 49.532 LbI Shunt
unused S/N 13368, 10.00 Lb , 3.00121 mV/v,
10.00 V, Cal on Oct27-99, 4.9292 Lb Shunt
unused S/N 63220, 100.00 Lb , 2.99984 mV/v,
10.00 V , Cal on Oct27-99, 49.381 Lb Shunt
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v,
5.00 V , Cal on Oct27-99, 486.45 Lb Shunt
Calibrate by Shunt Command (CS)
The Calibrate by Shunt (CS) command is used to calibrate the cell using the shunt
calibration factor expressed in pounds (or pound-inches for a torque cell). This
command will only function if you have already issued a Calibrate Begin (CB)
command. Note that the cell should not have any load or torque on it for this command
to operate correctly. The format is CS(shunt value in Lb or LbI)#. The example shows
calibration of a load cell using a shunt value of 553.26 Lb.
Example:
@123CS553.26#
Acknowledge:
@123 Calibrate Shunt Command - Reading...
[10 seconds delay]
@123 Calibrate Command - Reading for Shunt Check...
[10 seconds delay]
@123 Calibrate Command Completed
Ch A = S/N 123456, 1000.0 Lb , 4.50014 mV/v,
10.00 V , Cal on Oct27-99, 553.26 Lb Shunt
unused S/N 63220, 100.00 Lb , 2.99984 mV/v,
10.00 V , Cal on Oct27-99, 49.381 Lb Shunt
Ch B = S/N 89991, 1000.0 Lb , 4.50015 mV/v,
5.00 V , Cal on Oct27-99, 486.45 Lb Shunt
Calibrate Counts-Per-Inch Command (CI)
This command has two subcommands, a view command (CIV) and a set command
(CIS).
MODEL 9840 PG 42 PUB. 2856-16