Pressure Systems, Inc. Model 9046 User’s Manual
Page 22 www.PressureSystems.com
Examples:
Correct the offset in channel 1 with channel 1 at 100 degree C :
Command:
h0001 100.0
Response:
0.294167 (offset coefficient returned)
Manually set channel 1offset coefficient to 0.4:
Command:
v00100 0.4
Normal
:
A
Response:
Error: N08
Read channel 1 offset coefficient:
Command:
u00100
Normal
:
0.4 (offset coefficient = 0.4)
Response:
Error: N08
Store coefficients to FLASH PROM
Command:
w08
Response:
A