Pressure Systems, Inc. Model 9046 User’s Manual
Page 34 www.PressureSystems.com
5.3 Reset (Command ’B’)
Purpose: Instructs the module to reset internal operating parameters, and to set all internal
control variables to their default "reset" state (see description below). The current TCP/IP socket
connection will remain open. Execution after a power off/on cycle is optional (unnecessary).
Command "B"
’B’ is the operation code.
Response "A" or "Nxx"
’A’ for acknowledge; ’N’ for negative acknowledge,
error code xx.
Description: The module returns to the following "reset" states if this command is executed:
• The Offset calibration adjustment coefficient for each transducer is set to the last value
stored in flash memory.
• The Gain calibration adjustment coefficient for each transducer is set to the last value
stored in flash memory.
• Number of Samples for Data Averaging is set to last value stored in flash memory
(factory default = 64).
• Any autonomous host data delivery streams defined by ’c’ sub-commands are reset
(undefined).
Example:
Command module to reset "power-up" defaults:
"
B"
Read following response:
"
A"
Note
NUSS does not use this command normally. You may execute it
manually via the ’Execute Commands’ or ’Execute Script’ functions
of NUSS - though this should be avoided if NUSS is currently
operating the module via other NUSS functions.