Pressure Systems, Inc. Model 9046 User’s Manual
Page 47 www.PressureSystems.com
"A"
"
A"
"
A"
Example 2:
To similarly acquire data at "relative" rates (1, 2, and 4) using a periodic hardware trigger
(assumed to also cycle periodically at some practical (adjustable) rate), enter the commands:
"c 00 1 000F 0 1 7 0"
"c 00 2 00F0 0 2 7 0"
"c 00 3 FF00 0 4 7 0"
Read responses:
"
A"
"
A"
"
A"
Note
The type of data (also called the Data Group) delivered by default
for each specified channel (when the streams are started) is from
the Primary Transducer (i.e., Temperature/ Resistance) in E. U.,
unless sub-command "05" is subsequently executed (before the
stream is started) to select other Data Groups (and/or a UTR Alarm
State Prefix) in each stream. The "05" sub-command can also de-
select the default Primary Transducer (Temperature/Resistance)
data in E. U.