Pressure Systems 9046 Scanner User Manual


 
Pressure Systems, Inc. Model 9046 User’s Manual
Page 53 www.PressureSystems.com
5.7.4 Command ’c’- Sub-command Index 03: Clear Stream(s)
This sub-command is used to "undefine" any previously configured host stream in a module,
one at a time, or all together. The sub-command’s format is:
Command
c 03 st
‘c’ is the operation code
‘03’ is the sub-command index (‘ii’) for
configuration
st’ is the stream identifier character (1, 2, or 3 or
0=all streams)
NOTE: all parameters are separated by a space
character.
Response
A” or “Nxx”
‘A’ for acknowledge; ‘N’ for negative acknowledge,
error code xx.
Autonomous
Packet
none generated
Description: This sub-command clears (un-defines) the particular specified host stream (st=1-
3), or un-defines “all configured” host streams with a single command (
st=0). Once cleared, a
stream must be reconfigured before it can be started.
Example:
Stop all the streams configured previously. Then clear (un-define) only stream 3. Finally,
resume the remaining defined streams 1 and 2:
"
c 02 0"
"
c 03 3"
"
c 01 0"
Read response:
"
A"
"
A"
"
A"