Pressure Systems, Inc. Model 9046 User’s Manual
Page 56 www.PressureSystems.com
Description:
This sub-command returns the current configuration information for a particular stream.
Returned values are defined the same as the “contributing” parameters of the separate sub-
commands:
Configure a Host Delivery Stream (“00”), Select Protocol (“06”), and Select Data in
a Stream
(“05”). Note that the ‘num’ field represents the number of scans actually returned so
far (= last sequence number returned, or =0 if stream has not yet started.
Example:
Return configuration information for stream 1
"
c 04 1"
Read response:
"
1 FFFF 0 20 7 32000 1 7002 200.200.200.1 0010"
The above example shows that stream scans all 16 channels. Data is acquired using hardware
trigger with one (1) data scan acquired for every trigger events. Data is returned in binary format
7. 32000 scans have been returned so far. Data is sent using UDP protocol to port 7002 at IP
address 200.200.200.1. Temperature/Resistance EU data
only is returned for the requested
channels.
Note
NUSS only uses this sub-command (silently, behind the scenes) if
a user-written command script (executed by the NUSS ’Execute
Script’ function) starts a stream without having fully defined it (with
other ’c 00’ and ’c05’ commands) first.