EMC Information
CAUTION
The RadiForce series require special precautions regarding EMC and need to be installed, put
into service and used according to the following information.
Do not use any cables other than the cables that provided or specified by us.
Using other cables may cause the increase of emission or decrease of immunity.
Do not put any portable and mobile RF communications equipment close to the RadiForce
series. Doing so may affect the RadiForce series.
The RadiForce series should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the equipment or system should be observed to verify
normal operation in the configuration in which it will be used.
Guidance and manufacturer’s declaration - electromagnetic emissions
The RadiForce series are intended for use in the electromagnetic environment specified below.
The customer or the user of the RadiForce series should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions
EN55011
Group 1 The RadiForce series use RF energy only for its internal function.
Therefore, its RF emission are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
EN55011
Class B The RadiForce series are suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
EN61000-3-2
Class D
Voltage fluctuations /
flicker emissions
EN61000-3-3
Complies
Guidance and manufacturer’s declaration - electromagnetic immunity
The RadiForce series are intended for use in the electromagnetic environment specified below.
The customer or the user of the RadiForce series should assure that it is used in such an environment.
Immunity test EN60601
test level
Compliance level Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
EN61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient / burst
EN61000-4-4
±2kV
for power supply lines
±1kV
for input/output lines
±2kV
for power supply lines
±1kV
for input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
EN61000-4-5
±1kV differential mode
±2kV common mode
±1kV differential mode
±2kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
EN61000-4-11
<5% U
T
, for 0.5 cycle
(>95% dip in U
T
)
40% U
T
, for 5 cycles
(60% dip in U
T
)
70% U
T
, for 25 cycles
(30% dip in U
T
)
<5% U
T
, for 5s
(>95% dip for U
T
)
<5% U
T
, for 0.5 cycle
(>95% dip in U
T
)
40% U
T
, for 5 cycles
(60% dip in U
T
)
70% U
T
, for 25 cycles
(30% dip in U
T
)
<5% U
T
, for 5s
(>95% dip for U
T
)
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the RadiForce series requires
continued operation during
power mains interruptions, it is
recommended that RadiForce
series be powered from an
uninterruptible power supply or
battery.
Note: U
T
is the a.c. mains voltage prior to
application of the test level.
Power frequency
(50/60Hz)
magnetic field
EN61000-4-8
3A/m 3A/m The power frequency magnetic
field should be measured in the
intended installation location to
assure that it is sufficiently low.