National Instruments NI 4070/4072 6 1/2-Digit FlexDMM Graphics Tablet User Manual


 
NI 4070/4072 Calibration Procedure 26 ni.com
19. Call niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_2_WIRE_RES
Range = 100e6
Resolution = 100
20. Call niDMM_Read and store the result as the 100 MΩ range offset.
21. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_2_WIRE_RES
Range = 10e6
Resolution = 10
22. Call niDMM_Read and store the result as the 10 MΩ range offset.
23. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_2_WIRE_RES
Range = 1e6
Resolution = 1
24. Call niDMM_Read and store the result as the 1 MΩ range offset.
25. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_2_WIRE_RES
Range = 100e3
Resolution = 0.1
26. Call niDMM_ConfigureMultiPoint with the following parameters:
Trigger Count =
1
Sample Count = 4
Sample Trigger = NIDMM_VAL_IMMEDIATE
Sample Interval = –1
27. Call niDMM_ReadMultiPoint with the following parameters:
Maximum Time =
NIDMM_VAL_TIME_LIMIT_AUTO
Array Size = 4
Average the results by summing the returned reading array of the
function and dividing by the returned actual number of points.
Store the result as the 100 kΩ range offset.